Probe calibrator KHT 6000B
The KHT 6000B was specifically designed for high voltage dividers of up to 10,000 : 1. The positive rectangular output voltage can be adjusted in the range of 1000 ... 6000 V.
The voltage given to the output can be read out at a DVM.
Technical specification
|
|
|
|
|
Rectangular output voltage |
+ 1000..6000 V
switchable
|
|
|
Accuracy DC (± digit) |
± 1 % |
|
|
Rise time |
< 45 ns |
|
|
Overshoot |
< 2 % |
|
|
Repeat cycle |
1 Hz |
|
|
Pulse duration |
1 ms |
|
|
Trigger output |
15 V |
|
|
Trigger pulse duration |
10 µs |
|
|
Supply voltage |
100..230 V AC, 50..60 Hz
|
|
|
Dimensions W x H x D |
approx. 235 x 140 x 375 mm |
|
All dynamic values measured at oscilloscope accuracy of ± 2 %
|